Fluid path - connectors

BioClamp

Overview

BioClamp® - Plastic Tri-Clamp

BioClamp is a patented plastic union clamp, specifically designed to meet the needs of busy pharmaceutical and bioprocessing laboratories.

Features and benefits

  • Lot numbers molded in for full traceability
  • Molded from reinforced Nylon 66 USP Class VI, FDA
  • Reduced distortion on polymeric fittings when subjected to heat
  • Suitable for sterilization by autoclave and gamma irradiation
  • Tamper evident feature available on ½” – ¾” to 4" BioClamp

Technical summary

Max. operating pressure10 bar
Max. operating pressure145 psi
Shelf life5 years
CertificationADCF, ISO 14644-1 Class 7, ISO 9001, ISO10993, USP Class VI 
TC Size½-¾ TC - 8 TC in

Technical Summary table lists features available within the range. See ‘Technical specifications’ tab for model specific information.

Specifications

Technical specifications

BioClamp
Max. operating pressure10 bar
Max. operating pressure145 psi
Fluid temperature range 2 to 40 °C
Fluid temperature range 35 to 104 °F
Operating temperature range -4 to 40 °C
Operating temperature range 40 to 104 °F
Autoclave stability135 °C (275 °F) for 30 minutes 
Gamma stabilityMax 50kGy 
Shelf life5 years
CertificationADCF, ISO 14644-1 Class 7, ISO 9001, ISO10993, USP Class VI 
TC Size½-¾ TC - 8 TC in
Quantity5, 10, 20, 40, 50, 100 PCS

Materials of construction

MaterialGlass reinforced Nylon 66

Validation

About the BioClamp validation guides

BioClamp is manufactured in ISO 14644 class 7 cleanrooms within a facility operating an ISO 9001 quality management system.

Developed for the biotechnology and pharmaceutical industry, BioClamp is suitable for sterilization by autoclave and gamma irradiation.

 

Validation summary guide

The summary guide summarises the validation/qualification testing that has been carried out on BioClamp.

View BioClamp summary guide - 520.5kb

Request full validation guide

Thank you for requesting our validation documentation. We have received your request and will respond within 48 hours.

Disclaimer: You hereby agree the provided validation guide is for private use only to assist in the qualification of the products listed within the validation guide.

You also agree that you will not copy, redistribute, transmit, assign, sell, broadcast, rent, share, lend, modify, adapt, edit or transfer the validation guide to any other party without prior consent from WMFTS.

Disclaimer: You hereby agree the provided validation guide is for private use only to assist in the qualification of the products listed within the validation guide.

You also agree that you will not copy, redistribute, transmit, assign, sell, broadcast, rent, share, lend, modify, adapt, edit or transfer the validation guide to any other party without prior consent from WMFTS.

Literature

Brochures

  1. BioPure overview (EN).pdf
    PDF - 863.8kb

Datasheet

Other Resources

  1. 8 inch BioClamp technical drawing.pdf
    PDF - 4.1mb
  2. 1 inch - 1½ inch BioClamp technical drawing.pdf
    PDF - 2.5mb
  3. 2 inch BioClamp technical drawing.pdf
    PDF - 2.6mb
  4. 3 inch BioClamp technical drawing.pdf
    PDF - 2.8mb
  5. 4 inch BioClamp technical drawing.pdf
    PDF - 3mb
  6. 2½ inch BioClamp technical drawing.pdf
    PDF - 2.7mb
  7. 6 inch BioClamp technical drawing.pdf
    PDF - 9.6mb
  8. ½ inch - ¾ inch BioClamp technical drawing.pdf
    PDF - 3mb

Order Information

Product codes

BioClamp product codes

Part Number Size Pack size Max Operating Pressure Max Cumulative Gamma Dose Material
NG075WHT 1/2" - 3/4" 10 150 psiG 10 bar 40 kGy Nylon 66
NG075WHT-100 1/2" - 3/4" 100 150 psiG 10 bar 40 kGy Nylon 66
NG150WHT 1" - 1 1/2" 10 150 psiG 10 bar 40 kGy Nylon 66
NG150WHT-50 1" - 1 1/2" 50 150 psiG 10 bar 40 kGy Nylon 66
NG200WHT 2" 10 150 psiG 10 bar 40 kGy Nylon 66
NG200WHT-50 2" 50 150 psiG 10 bar 40 kGy Nylon 66
NG250WHT 2 1/2" 10 150 psiG 10 bar 40 kGy Nylon 66
NG250WHT-40 2 1/2" 40 150 psiG 10 bar 40 kGy Nylon 66
NG300WHT 3" 10 150 psiG 10 bar 40 kGy Nylon 66
NG300WHT-20 3" 20 150 psiG 10 bar 40 kGy Nylon 66
NG400WHT 4" 5 150 psiG 10 bar 40 kGy Nylon 66
NG400WHT-20 4" 20 150 psiG 10 bar 40 kGy Nylon 66
NG600WH 6" 5 90 psiG 6 bar 40 kGy Nylon 66
NG800WH 8" 5 90 psiG 6 bar 40 kGy Nylon 66

Recommended Products

  1. Pumpsil tubing

    Pumpsil tubing

    Platinum-cured silicone tubing with ultra smooth bore and excellent flow stability. -4 ºF to 176 ºF (-20 ºC to 80 ºC).

    Tubing - peristaltic

  2. PureWeld XL tubing

    PureWeld XL tubing

    High performance TPE tubing with exceptional life and accurate flow rates. Weldable and sealable for sterile connectivity.

    Tubing - peristaltic

Related Products

Our specialists are here to help you succeed.

To order a product or ask a question

Contact our experts